Impulse
Noise
Count
CAUTION
Ensure
that
the
AC
Calibrator
OUTPUT
is
switched
off
or
set
to
the
minimum
level
before
proceeding
with
the
following
procedure.
1
Connect
the
AC
Calibrator
output
to
the
HP
4934A
RCV
port.
2
Set
the
HP
4934A
as
follows:
Measurement
:
IMPULSE
NOISE
Transmitter
(TRMT)
Impedance
:
600Q
Receiver
(RCV)
Impedance
:
600Q
Display
Mode
:
RCV
Filter
:15
kHz
FLAT
Period
:
NON-STOP
Impulse
Noise
Display
:
LOW
(SET)
3
Set
the
AC
Calibrator
output
level
to
0.7746
V
at
2,000
Hz
then
switch
on
the
OUTPUT.
4
Set
the
LOW
(SET)
threshold
to
75
dBm.
5
Press
START
RESET
to
start
the
measurement.
6
Wait
60
seconds
then
press
STOP.
7
Check
the
reading
on
the
HP
4934A
display
is
480
±
48
counts.
8
Select
the
MID
threshold.
Check
the
reading
on
the
display
is
480
±
48
counts.
9
Select
the
HIGH
threshold.
Check
the
reading
on
the
display
is
480
±
48
counts.
Performance
Tests
5-9