Transmitter
Flatness
at
-40
dBm
SPECIFICATION
Transmitter
Flatness
20
Hz
to
200
Hz
:
±1.0
dB
200
Hz
to
15
kHz
:±0.2dB
15
kHz
to
85
kHz
:±0.5dB
DESCRIPTION
An
AC
Voltmeter,
AC
Calibrator
and
20
dB
Amplifier
are
used
to
check
the
HP
4934A
transmitter
flatness
at
-40
dBm
over
the
frequency
range
20
Hz
to
85
kHz.
The
20
dB
Amplifier
is
required
to
increase
the
-40
dBm
level
to
within
the
specified
range
of
the
AC
Voltmeter.
The
amplifier
provides
a
600Q
termina¬
tion
and
is
powered
from
a
separate
DC
Power
Supply.
If
a
high
input
impedance
amplifier
is
used,
a
600Q
termination
is
required.
Variations
in
the
gain
of
the
20
dB
Amplifier
are
compensated
for
at
each
of
the
test
frequencies
in
the
following
procedure.
EQUIPMENT
AC
Voltmeter
AC
Calibrator
20
dB
Amplifier
Dual
Output
DC
Power
Supply
PROCEDURE
CAUTION
Ensure
that
the
AC
Calibrator
OUTPUT
is
switched
off
or
set
to
the
minimum
level
before
proceeding
with
the
following
procedure.
:
HP
3456A
:
DATRON
4707A
:
SAl
(see
Appendix
A)
:
HP
6234A
5-24
Performance
Tests