Hardware
Error
Messages
Local: These messages
are
all
queued
up
and
can
be displayed using
the
[
MSSG
) key.
HP-IB:
All
of
these messages are
output
via
HP-IB
using "OH". (Some
out-of-Iock
messages
are
also
output
using "0H";
refer
to page
3-197.)
Hardware
Error
messages (message
number
is
prefixed
with
"H") signal
the
presence
of
either
an
internal
circuit
hardware
problem
or
a
firmware
problem
that
you
might
be
able
to
resolve.
Hardware
Error
messages
may
occur
independently
of
function
execution. A
maximum
of
20
Hardware
Error
messages
can
be
queued
up
to
be accessed
either
by
the
[
MSSG
)
key
on
the
front
panel,
or
via
HP-IB
using "OH".
If
you
are
unable
to
repair
the
hardware
error,
refer
to
the
Service
Manual
for
more
information.
HIT
AMPTD=CLEAR
RVS.PWR
.Hl
7001
Press
the
[
AMPTD
)
key
to
reset
the
reverse-power-protect
circuitry.
A
reverse-power
condition
has
been
sensed.
After
this
condition
is
eliminated,
select
the
(
AMPTD
I
kflY
to reset
the
protection
circuitry.
HP-IB:
Send
AP
after
the
condition
is
eliminated.
Message
H2
will
always
occur
after
the
reverse
-power-protect
circuitry
is
reset
in
response
to
message
HI.
R VS.PWR
PROTECT
CLEARED
.H2
7002
The
reverse-power
protect
circuitry
was
previously
tripped
but
is
now reset.
RECALL
ERROR
FOUND
.HI0
7010
A
recall
error
has
been
found.
This
message
could
be
due
to
a
hardware
failure
(for example, loss
of
RAM
memory).
The
message
means
that
a
recall
register
that
was
recalled
had
no
instrument
settings
saved
in
it.
The
register
could
have
been
cleared
with
Special
Function
251 (Clear
Recall
Registers)
or
because
of
a
hardware
error.
This
error
HI
0 occurs
in
conjunction
with
execution
error
E93.
Refer
to
"Recalling
an
Undefined
Register"
on
page 3
-162
for
more
information.
MSSG
BUFFER
OVERFLOWED
.Hll
7011
The
message
buffer
has
overflowed. Some messages were lost.
lOMHZ
REF.
OVEN
COLD
.H20
7020
The
oven
for
the
10
MHz
reference
(installed
with
option
001)
is
cold.
The
frequency
reference
could
be
unstable
until
the
oven
is
warm.
When
the
oven
cold
condition
no
longer
exists,
an
attempt
to scroll
through
the
message list
twice
will
not
display
the
message
the
second time.
(Refer
to
"Queued
Up
Messages"
on
page
3-180
for
an
explanation
of
the
purpose
of
scrolling
through
the
message list twice.)
Also
note
that
when
this
message
is
available,
the
Knob
might
not
operate
in
the
normal
way.
3-193

Table of Contents

Other manuals for HP 8642A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8642A and is the answer not in the manual?

HP 8642A Specifications

General IconGeneral
BrandHP
Model8642A
CategoryInverter
LanguageEnglish

Related product manuals