12
iM3 Dental Systems
Operation Recommendations
Air pressure The dental unit air pressure
should be adjusted to be
tween 35-50 psi (normally 40
psi). See page 19
Lubrication To assure optimal
performance and life the
handpiece should be
lubricated daily with Triflow
®
Teflon lubricant. Instructions
for lubrication are found on page
20.
Air supply Moist or dirty air from the
compressor can cause damage.
Check the regulator filter regu-
larly. If dirt or moisture is pre-
sent, replace the regulator filter
and bleed the system, to remove
water or oil.
Installation Instructions
1. Preset the air pressure between 35 and 45
psi, normally 40 psi (see page 19)
2. Connect the low speed handpiece to the
grey handpiece line
3. Fit the 4:1 green straight nose cone. Attach
the Blue R&R oscillating prophy head
4. Depress the foot control fully to operate the
low speed handpiece
Note: Polishing should be carried out at 5,000 rpm or less. This can be
accomplished by running the motor at full revs (20,000 Rpm) and use a 4:1 re-
duction head (green ), to achieve 5,000 revs, maintaining torque.
iM3 Advantage Low Speed Handpiece
The low speed handpiece features include a light weight balanced profile and
smooth swivel action. Its vane type power system is quiet and virtually vibration
free. The motor accommodates both doriot or ISO heads and its powerful torque is
suitable for all low speed procedures.
iM3 low speed handpieces feature a 5 blade turbine for greater torque, not found on
other LS handpieces. iM3 5BT