Handpiece Trouble Shooting
Low Speed Handpiece
PROBLEM CAUSE SOLUTION
Handpiece
won’t run
1. No or low air pressure.
2. Air line improperly sealed.
3. Inferior oil; too much or excess
more commonly too little oil or
water in the turbine.
4. Broken drive head.
Note new handpieces require a
“Run in time” this can be up to 20
hours of use
1. Check the air line to be sure
the air is turned on and the
line is unobstructed.
2. Reconnect air line to
handpiece making sure the
gasket is fully contracted.
3. See lubrication instructions
on page 19.
4. Contact your local official
Service Centre.
If the motor stops spinning
place a fine screw driver down
the nose and rotate shaft.
Low Speed Handpiece - Straight Nose Cone
PROBLEM CAUSE SOLUTION
Chuck Difficult to
Open or Close
1. Chuck housing screw missing.
2. Improper lubrication.
1. Replace chuck housing
screw, contact your local
official iM3 Service Centre.
2. See lubrication instruc-
tions, on page 19.
iM3 Inc iM3 Pty Ltd
12119 NE 99th Street Suite 2060 9/31-33 Chaplin Drive
Vancouver WA 98682 USA Lane Cove NSW 2066 Australia
Tel (360) 254 2981 Fax (360) 254 2940 Tel (02) 9420 5766 Fax (02) 9420 5677
www.im3vet.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the iM3 GS and is the answer not in the manual?

iM3 GS Specifications

General IconGeneral
BrandiM3
ModelGS
CategoryDental equipment
LanguageEnglish