20
iM3 Dental Systems
Sterilization
HIGH SPEED HANDPIECE
Handpieces should be thoroughly cleaned to remove any foreign particles, eg; blood, saliva
etc. Make sure that the turbine cartridge and head cavity are clean before autoclaving. If
autoclaved with bloody particles in the turbine housing, autoclaving will cause particles to
be scorched and damage to the turbine may result.
The exterior of the handpiece may be cleaned with any good surgical disinfectant. Be sure
to rinse and dry thoroughly. Lubricate handpiece turbine with Triflow Teflon Lubricant.
Note: The neoprene washer or connections gasket should be removed prior to autoclaving.
After autoclaving, re-lubricate when handpiece is cool.
Do not allow handpiece to sit in autoclave bag overnight.
LOW SPEED HANDPIECE
Prepare handpiece and attachments for sterilization by carefully cleaning the exterior
surface. Thoroughly scrub, rinse and clean away any residual solution and particles.
Remove moisture with a towel or use air from the 3 way air water syringe.
1. Prior to sterilization, lubricate the handpiece using Triflow Teflon Lubricant.
HANDPIECES: Apply oil through the handpiece air inlet tube. (see page 19)
ATTACHMENT: Remove attachment from handpiece and oil through both end openings.
Note: After lubricating items, assemble and operate for a few seconds to expel excess lubricant.
2. Seal items in sterilization tubing bags. Make sure to remove bur from handpiece prior to sealing.
3. Place handpiece or attachments in sterilization trays of dry heat sterilizer or steam autoclave. Do
not exceed temperatures of 275°F or 135°C. Sterilize according to manufacturers instructions.
4. After sterilization cycle is completed, allow handpieces and attachments to cool down then
lubricate handpiece and attachments thoroughly. Make sure to use a different container after
sterilization cycle, to prevent recontaminating.
Note: Lubricate handpieces using Triflow® Teflon Lubricant after cooling down, do not
allow handpiece to sit in sealing bag overnight or for a long period. Immediately operate
handpiece as described in the following step.
5. After lubricating items, insert bur into handpieces and operate for a few seconds to
remove excess lubricant.
Do not autoclave the high speed handpiece with a bur in the chuck.