Status Commands
Copyright © Itech Electronic Co., Ltd. 3
Chapter3 Status Commands
STATus:OPERation[:EVENt]?
This command can read the parameter from operation event register.
After executing this order, operation event register is reset.
Syntax
STATus:OPERation[:EVENt]?
Arguments
None
Example
STAT:OPER?
Returns
NR1
STATus:OPERation:CONDition?
This command can read the parameter from the operation condition register.
When the parameter of operation condition register changes, the bit
corresponding in operation event register is 1.
Syntax
STATus:OPERation:CONDition?
Arguments
None
Example
STAT:OPER:COND?
Returns
NR1
STATus:OPERation:ENABle <n>
This command can set the parameter of operation event enable register.
Setting parameter can determine which bit value of operation event register is 1
and the bit will cause OPER of status byte register to be 1.