System-Related Functions
• Re-Energy: indicates the regenerative power.
• AHour: indicates the capacity of charge and discharge, when the battery
simulator function or battery test function is enabled, the capacity data
will be cleared.
• WHour: indicates energy data, when the battery simulator function or
battery test function is enabled, the energy data will be cleared.
• Reset Meas: clearing the energy data and capacity data.
• Reset ReE: clearing the regenerative power data.
4. Press [Esc] to exit.
7.7 External analog function (optional)
When the interface card selected by the user is DB25 analog interface (IT-
E1208), the analog interface can realize the external analog function.
The rear panel of this series is optionally configured with a DB25 analog inter-
face, through which, the following functions can be realized:
• Remotely control voltage and current values
• Control On/Off state
• Remotely monitor CC and CV states as well as voltage and current values
• Clear instrument faults
• Monitor the instrument loading state, On/Off state and fault state
• RS-485 interface connection. For detailed introduction about interface, refer
to 2.6.5 RS-485 Interface
• Before connecting the hardware equipment that controls the analog in-
terface, ensure that the hardware equipment would not output voltage
that is 20% higher than specified value. Otherwise, it would damage the
instrument. For example, during setting of voltage or current, the input
voltage cannot exceed 12V. Otherwise, it would damage the instrument.
• Under the analog remote control mode, the analog signal input pin can-
not be suspended, before enabling the external analog function on the
front panel, it is necessary to correctly connect the 1 (Vol_Pro), 15 (Cur_
Pro) and 17 (Cur/Vol_Pro).
• There is safe electrical isolation between the analog interface and the
DC terminal. Do not connect any ground wire of the analog interface to
the DC+ or DC- terminal on the instrument’s rear panel.
Copyright © Itech Electronic Co., Ltd.
83