EasyManuals Logo
Home>Keithley>Multimeter>2701

Keithley 2701 Service Manual

Keithley 2701
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #150 background imageLoading...
Page #150 background image
Model 2701 Service Manual Calibration Reference B-11
:STEP5
(:CALibration:PROTected:DC:STEP5)
Purpose To program the 100V DC comprehensive calibration step.
Format :cal:prot:dc:step5 <Cal_voltage>
Parameter <Cal_voltage> = 90 to 110 [V]
Description :STEP5 programs the 100V DC comprehensive calibration step. The
allowable range of the calibration voltage parameter is from 90 to 110, but
100 is recommended for best results.
Example :CAL:PROT:DC:STEP5 100 Program 100V step.
:STEP6
(:CALibration:PROTected:DC:STEP6)
Purpose To program the 1k 4-wire comprehensive calibration step.
Format :cal:prot:dc:step6 <Cal_resistance>
Parameter <Cal_resistance> = 900 to 1.1E3 []
Description :STEP6 programs the 1k 4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from 900
to 1.1E3, but 1E3 is recommended for best results.
Example :CAL:PROT:DC:STEP6 1E3 Program 1k step.
:STEP7
(:CALibration:PROTected:DC:STEP7)
Purpose To program the 10k 4-wire comprehensive calibration step.
Format :cal:prot:dc:step7 <Cal_resistance>
Parameter <Cal_resistance> = 9E3 to 11E3 []
Description :STEP7 programs the 10k 4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from
9E3 to 11E3, but 10E3 is recommended for best results.
Example :CAL:PROT:DC:STEP7 10E3 Program 10k step.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals