Model 2701 Service Manual Calibration 2-17
AC voltage calibration
Follow these steps for AC voltage calibration:
1. Connect the calibrator to the Model 2701 INPUT HI and LO terminals as shown in
Figure 2-4.
2. Perform the calibration steps summarized in Table 2-11. For each step:
Set the calibrator to the indicated voltage and frequency, make sure the unit is in
operate. (You must use the stated voltage and frequency.)
Send the indicated programming command.
•Wait until the Model 2701 completes each step before continuing.
Table 2-11
AC voltage calibration programming steps
Calibration step Calibrator voltage, frequency Calibration command
10mV AC at 1kHz
100mV AC at 1kHz
100mV AC at 50kHz
1VAC at 1kHz
1VAC at 50kHz
10VAC at 1kHz
10VAC at 50kHz
100VAC at 1kHz
100VAC at 50kHz
700VAC at 1kHz
10.00000mV, 1kHz
100.0000mV, 1kHZ
100.0000mV, 50kHz
1.000000V, 1kHz
1.000000V, 50kHz
10.00000V, 1kHz
10.00000V, 50kHz
100.0000V, 1kHz
100.0000V, 50kHz
700.000V, 1kHz
:CAL:PROT:AC:STEP1
:CAL:PROT:AC:STEP2
:CAL:PROT:AC:STEP3
:CAL:PROT:AC:STEP4
:CAL:PROT:AC:STEP5
:CAL:PROT:AC:STEP6
:CAL:PROT:AC:STEP7
:CAL:PROT:AC:STEP8
:CAL:PROT:AC:STEP9
:CAL:PROT:AC:STEP10

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals