B-18 Calibration Reference Model 2701 Service Manual
:CODE
(:CALibration:PROTected:CODE)
Purpose To program the calibration code or password so that you can perform the
Model 7700 calibration procedures.
Format :cal:prot:code '<char_string>'
Parameter Up to a 8-character string including letters and numbers.
Description The :CODE command enables the Model 2701 calibration procedures when
performing these procedures over the bus. This command must be sent to
the unit before sending any other Model 7700 calibration command. The
default calibration code is KI002701.
NOTES The
:CODE
command should be sent only once before performing calibration. Do
not send
:CODE
before each calibration step.
The code parameter must be enclosed in single quotes.
Example :CAL:PROT:CODE 'KI002701' Send default code of KI002701.
:COUNt?
(:CALibration:PROTected:CARD1:COUNt?)
(:CALibration:PROTected:CARD2:COUNt?)
Purpose To determine how many times a Model 7700 has been calibrated.
Format :cal:prot:card1:coun?
:cal:prot:card2:coun?
Response <n> Calibration count.
Description The :CARD1:COUNt? and :CARD2:COUNt? commands allow you to
determine how many times a Model 7700 in Slot 1 and Slot 2 respectively
has been calibrated.
NOTE Use the
:COUNt?
command to help you monitor for unauthorized calibration
procedures.
Example :CAL:PROT:CARD1:COUN? Request card 1 calibration count.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals