Model 2701 Service Manual Performance Verification 1-21
2. Install the Model 7700 in Slot 1 of the Model 2701, then turn on the power and allow
the unit to warm up for two hours before proceeding. Be sure the front panel INPUTS
switch is set to the REAR position.
3. Select the AC volts function by pressing the ACV key. Close Channel 1 by pressing the
CLOSE key and then keying in 101.
4. Set the Model 2701 for the 100mV range, make sure that REL is disabled.
5. Source 1kHz and 50kHz AC voltages for each of the ranges summarized in Table 1-10
and make sure that the respective Model 2701 readings fall within stated limits.
6. Press the OPEN key to open Channel 1.
Table 1-10
Plug-in module ACV reading limits
ACV
range
Applied AC
voltage
1kHz reading limits
(1 year, 18°C to 28°C)
50kHz reading limits
(1 year, 18°C to 28°C)
100mV
1V
10V
100V
750V
100.0000mV
1.000000V
10.00000V
100.0000V
300.000V*
99.910 to 100.090mV
0.99910 to 1.00090V
9.9910 to 10.0090V
99.910 to 100.090V
299.60 to 300.40V
99.830 to 100.170mV
0.99830 to 1.00170V
9.98300 to 10.0170V
99.830 to 100.170V
299.27 to 300.73V
* If the 5725A amplifier is not available, change the 300V @ 50kHz step to 220V @ 50kHz. Read-
ing limits for 220V @ 50kHz = 219.36 to 220.64V.

Table of Contents

Other manuals for Keithley 2701

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2701 and is the answer not in the manual?

Keithley 2701 Specifications

General IconGeneral
BrandKeithley
Model2701
CategoryMultimeter
LanguageEnglish

Related product manuals