EasyManuals Logo
Home>Keithley>Switch>2790

Keithley 2790 User Manual

Keithley 2790
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #29 background imageLoading...
Page #29 background image
2-8 Model 2790 SourceMeter
®
Switch System User’s Manual
Remote programming
As shown in Table 2-1, the amplitude command is used to set the outputs of the sources.
Notice that the channel number of the V-source (28) and I-source (27) must be included in
the command string.
Table 2-1
Commands to set amplitude for V-source and I-source
Commands Description Default
SOURce:VOLTage[:LEVel][:IMMediate]
[:AMPLitude] <NRf>, <clist> Set V-source output; 50 to 500 (V).
<clist> = (@128) Slot 1
= (@228) Slot 2
50V
[:AMPLitude]? <clist> Query V-source output level.
SOURce:CURRent[:LEVel][:IMMediate]
[:AMPLitude] <NRf>, <clist> Set I-source output; 0 to 50 (mA).
<clist> = (@127) Slot 1
= (@227) Slot 2
0.001A
[:AMPLitude]? <clist> Query I-source output level.
Note: Selecting an ohms calculation enables source tracking. Selecting any other math function disables
source tracking.