Model 2790 SourceMeter
®
Switch System User’s Manual 5-21
Table 5-4
Test procedure – contact checks for dual stage inflator
Test Step Command Sequence Front Panel Operation
1) Reset 2790 and open all
channels.
*RST Press SHIFT > press SETUP > select RESTORE:
FACT.
2) Select 2. SENS:FUNC ‘RES’ Press 2.
3) Close channels to test HI
contact for bridgewire A.
ROUT:MULT:CLOS
(@101,114,118)
Press CLOSE > select MULTI > select
101 > select 114 > select 118 > press EXIT.
4) Verify channels that are
closed (optional).
ROUT:MULT:CLOS? Press SHIFT > press CARD > select VIEW > select
SLOT1: 7751 or 7753. Press EXIT when finished.
5) Perform measurement. READ? Take reading from display.
6) Open all channels. ROUT:OPEN:ALL Press OPEN > select ALL.
7) Close channels to test HI
contact for bridgewire B.
ROUT:MULT:CLOS
(@104,114,118)
Press CLOSE > select MULTI > select
104 > select 114 > select 118 > press EXIT.
8) Verify channels that are
closed (optional).
ROUT:MULT:CLOS? Press SHIFT > press CARD > select VIEW > select
SLOT1: 7751 or 7753. Press EXIT when finished.
9) Perform measurement. READ? Take reading from display.
10) Open all channels. ROUT:OPEN:ALL Press OPEN > select ALL.
11) Close channels to test
housing contact.
ROUT:MULT:CLOS
(@108,115,118)
Press CLOSE > select MULTI > select
108> select 115 > select 118 > press EXIT.
12) Verify channels that are
closed (optional).
ROUT:MULT:CLOS? Press SHIFT > press CARD > select VIEW > select
SLOT1: 7751, 7752, or 7753. Press EXIT when
finished.
13) Perform measurement. READ? Take reading from display. Reading should be 2 or
less.
14) Open all channels. ROUT:OPEN:ALL Press OPEN > select ALL.
Test step notes:
1. Resetting the 2790 opens all channels and selects the DCV function (auto range). For remote programming, *RST places the
2790 in the one-shot trigger mode. For front panel operation, FACT defaults places the 2790 in the continuous trigger mode.
2. This step selects the 2 measurement function of the Model 2790.
3. This step closes the channels to make the connections for the HI contact test (bridgewire A).
4. This step is optional. Use it to verify that the correct channels are closed for the test.
5. For remote programming, READ? triggers and requests a single reading.
6. Always open all channels at the beginning and end of a test.
7. This step closes the channels to make the connections for the HI contact test (bridgewire B).
8. This step is optional. Use it to verify that the correct channels are closed for the test.
9. For remote programming, READ? triggers and requests a single reading.
10. Always open all channels at the beginning and end of a test.
11. This step closes the channels to make the connections for the housing contact test.
12. This step is optional. Use it to verify that the correct channels are closed for the test.
13. For remote programming, READ? triggers and requests a single reading.
14. Always open all channels at the beginning and end of a test.