Section 2: Performance verification Model DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
2-30 DMM6500-905-01 Rev. A/July 2018
Verify digitize current 100 mA range
Verify digitize current 1 A range
Verify digitize current 3 A range
* The 3 A range full-scale test points are limited to 2 A in this table because of the accuracy
limitations of Fluke Models 57xxA and 5725A series calibrators at currents above 2.2 A.
Verify digitize current 10 A range
See Digitize current 10 A range verification (on page 2-39) under Rear-panel verification (on page 2-
37).
AC current verification
The following topics describe how to verify AC current.
Verify AC current on the 100 µA to 3 A ranges
To verify AC current accuracy, you will:
• Apply accurate voltages from the Fluke 5720A or 5730A multifunction calibrator to the DMM6500
front-panel terminals.
• Verify that the displayed readings fall within specified limits.