Model 2450
Interactive SourceMeterĀ® Instrument User's Manual Section 3:
Using a remote interface
2450-900-01 Rev. C / December 2013 3-3
Figure 9: IEEE-488 connection example
Set the GPIB address
The factory sets the GPIB address value to 18. You can set the address to any address from 0 to 30
if it is unique in the system. This address cannot conflict with an address that is assigned to another
instrument or to the GPIB controller.
GPIB controllers are usually set to 0 or 21. To be safe, do not configure any instrument to have an
address of 0 or 21. To change the controller address, see the documentation for the controller.
The instrument saves the address in nonvolatile memory. This means that it does not change when
you send a reset command or when you turn the power off and then on again.
To set the GPIB address from the front panel:
1. Press the MENU key.
2. Under System, select Communication. The SYSTEM COMMUNICATION window opens.
3. Select the GPIB tab.
4. Next to Address, select the number. The Set GPIB Address dialog box is displayed.
5. Enter the address.
6. Select OK.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley Interactive SourceMeter 2450 and is the answer not in the manual?

Keithley Interactive SourceMeter 2450 Specifications

General IconGeneral
BrandKeithley
ModelInteractive SourceMeter 2450
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals