Section
6: Leakage current and insulation resistance Model 2450 Interactive SourceMeter® Instrument
User's Manual
6-4 2450-900-01 Rev. C / December 2013
Measuring leakage current
The following application demonstrates how to use the Model 2450 to measure the leakage current of
a 1 nF capacitor by sourcing a voltage and measuring the resulting current using the front panel or
over the remote interface. The remote interface examples show SCPI commands and TSP
commands.
This application sets the Model 2450 to source 20 V and measure the resulting leakage current as a
function of time. The instrument takes current measurements for a specific period.
For this test, you will:
Reset the instrument
Set the instrument to read the rear terminals
Select the source voltage function and measure current function
Set the magnitude of the voltage source
Turn on autoranging
Set the measure delay
Use the Duration Loop trigger model template to make readings for a specified time period
Turn on the source output
Take readings for a specified period of time
Turn off the source output
When you use the Model 2450 to measure small current values, ensure that the device under test is
electrostatically shielded. If the capacitor rating is greater than 20 nF, enable the high capacitance
mode for best results.
For more information about making optimized capacitor leakage measurements and minimizing
noisy measurements, see the Keithley Instruments Low Level Measurements Handbook, available
on the Keithley Instruments website (http://www.keithley.com).
Set up the leakage current application using the front panel
To set up the application from the front panel:
1. Use the test leads to make the connection from the capacitor to the rear panel of the Model 2450,
as described in Device connections (on page 6-2
).
2. Reset the Model 2450.
a. Press the MENU key.
b. Under System, select Manage.
c. Select System Reset, and then select OK.
3. Press the TERMINAL FRONT/REAR switch to set the instrument to use the rear-panel terminals
(R is displayed to the left of the switch).
4. Press the HOME key.
5. Press the FUNCTION key.
6. Under Source Voltage and Measure, select Current.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley Interactive SourceMeter 2450 and is the answer not in the manual?

Keithley Interactive SourceMeter 2450 Specifications

General IconGeneral
BrandKeithley
ModelInteractive SourceMeter 2450
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals