S530 Diagnostic Manual Section 1: Introduction to S530 Diagnostics
Figure 9: Blank probe card: Keithley 9139-type
SMU diagnostic testing
The first test that will execute is the Diag_SMU test, which identifies and then runs all the tests
required for the Series 2600A instruments and Model 2410 System SourceMeter
®
instruments.
NOTE
The purpose of SMU diagnostic is to check the functionality of Series 2600A instruments and 2410
instruments, not to verify the accuracy of the output and measurement.
The Setup tab for the SMU diagnostic contains the following (see next Figure):
KI26XX SMU and KI24XX SMU: In this area, all scanned SMUs listed in the Hardware Configuration
panel are displayed. Each SMU can be selected (or deselected) for testing.
Two functions are displayed (see next Figure):
SELECT ALL: Allows you to select all the SMUs. You can also manually select the individual SMUs.
CLEAR ALL: Deselects all of the SMUs.
SELECT ITEM: Allows you to select or deselect individual diagnostic tests. There are three kinds:
Communication, Voltage, and Current.
LOOP: The number entered is the number of times the selected set of tests will be repeated when
Diag_SMU is executed. The tests can be repeated once to 99 times.
S530-906-01 Rev. A / March 2011 1-11

Table of Contents

Other manuals for Keithley S530

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley S530 and is the answer not in the manual?

Keithley S530 Specifications

General IconGeneral
BrandKeithley
ModelS530
CategoryTest Equipment
LanguageEnglish

Related product manuals