EasyManuals Logo
Home>Keithley>Test Equipment>S530

Keithley S530 User Manual

Keithley S530
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #114 background imageLoading...
Page #114 background image
Section
3: LPTLib command reference S530/S540 KTE Linear Parametric Test Library (LPTLib) User's Manual
3-92 S530-900-01 Rev. E / September 2017
Example 1
double res22, vcc8;
.
.
conpin(SMU1, 3, 0);
conpin(SMU2, 2, 0);
conpin(GND, 1, 0);
forcev(SMU2, vcc8); /* Apply collector voltage to vcc8. */
trigig(SMU2, +5.0E-3); /* Search for a collector */
/* current of 5 mA. */
searchi(SMU1, 5.0E-5, 2.0E-4, 15, 1.0E-3, &res22); /* Generate */
/* a current ranging */
/* from 50 uA to 200 uA in */
/* 15 iterations. Return the */
/* current resulting from the */
/* last iteration as res22. */
This example uses the trigig and searchi commands together to generate and search for a specific
current level. A search is initiated to find the base current needed to produce 5 mA of collector current. The
collector-emitter voltage supplied by SMU2 is defined by the variable vcc8. The searchi command
generates the base current from SMU1. This current ranges between 50 mA and 200 mA in 15 iterations.
The trigig command continuously monitors the current through SMU1. The base current supplied by
SMU1 is stored as the result res22.

Table of Contents

Other manuals for Keithley S530

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley S530 and is the answer not in the manual?

Keithley S530 Specifications

General IconGeneral
BrandKeithley
ModelS530
CategoryTest Equipment
LanguageEnglish

Related product manuals