3: LPTLib command reference S530/S540 KTE Linear Parametric Test Library (LPTLib) User's Manual
3-36 S530-900-01 Rev. E / September 2017
kibdefclr
This command defines the device-dependent command sent to an instrument connected to the GPIB1 interface.
Usage
int kibdefclr(int pri_addr, int sec_addr, unsigned int timeout, double delay, unsigned
int snd_size, char *sndbuffer);
pri_addr
The primary address of the instrument (0 to 30; the controller uses address 31)
sec_addr
The secondary address of the instrument (1 to 30; if the instrument device does not
support secondary addressing, this parameter must be -1)
timeout
The GPIB timeout for the transfer in 100 ms units (for example,
timeout = 40 = 4.0 s)
delay
The time to wait after the device-dependent string is sent to the device, in seconds
snd_size
The number of bytes to send over the GPIB interface
sndbuffer
The physical byte buffer containing the data to send over the bus (the physical
CLEAR string); a maximum of 1024 bytes is allowed
Details
This string is sent during any normal tester-based devclr command. It ensures that if the tester is
calling the devclr command internally, any external GPIB device is cleared with the given string.
Each call to the kibdefclr and kibdefint commands copies parameters into a data structur
e
w
ithin the tester memory. These data structures are allocated dynamically. These tables are cleared
when the devint command executes. Any strings previously defined must be redefined.
The tester system allows you to define a maximum of 20 clear and 20 initialization strings. Each string
may contain up to a maximum of 1024 bytes.
Strings are sent over the GPIB interface in a first-in, first-out queue. This means that the first call to
the kibdefclr or kibdefint command is the first string sent over the GPIB. The devclr
(kibdefclr) strings are always sent before initialization.
The KIBLIB devclr strings are sent before the devclr and devint commands execute. This may
be a problem when communicating with any Keithley-supported GPIB instruments. This may also
have an effect on the bsweepX command, because the bsweepX command sends a call to t
he
devclr
command to clear active sources. It is not recommended to use GPIB instruments whe
n
per
forming tests with the bsweepX command.
Also see
bsweepX (on page 3-15)
devclr (on page 3-26)
devint (on page 3-26)
kibdefint (on page 3-37)

Table of Contents

Other manuals for Keithley S530

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley S530 and is the answer not in the manual?

Keithley S530 Specifications

General IconGeneral
BrandKeithley
ModelS530
CategoryTest Equipment
LanguageEnglish

Related product manuals