4: Error definitions S530/S540 KTE Linear Parametric Test Library (LPTLib) User's Manual
4-6 S530-900-01 Rev. E / September 2017
650 TAPI_BADCHANNEL
Message: Request to open unknown channel type XXX.
Remarks: An internal system software error has occurred.
651 TAPI_BADTESTER
Message: **This is a host-side only error. No message will be generated by the tester.
Remarks: There is no network node corresponding to the tester address. Make sure the network
configuration is correct.
652 TAPI_NOTFOUND
Message: **This is a host-side only error. No message will be generated by the tester.
Remarks: The tester cannot be located on the network. Make sure the tester is powered on and has
started correctly. This problem can also be caused by network problems.
653 TAPI_REFUSED
Message: **This is a host-side only error. No message will be generated by the tester.
Remarks: The tester is in use. This can be caused by another process running diagnostics or a test
plan.
656 TAPI_CHANLIMIT
Message: Channel limit exceeded.
Remarks: There are too many active network connections to the tester. Close some of the tools that
communicate with the tester and try again.
657 TAPI_BUFOFLOW
Message: **This is a host-side only error. No message will be generated by the tester.
Remarks: An internal system software error has occurred.

Table of Contents

Other manuals for Keithley S530

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley S530 and is the answer not in the manual?

Keithley S530 Specifications

General IconGeneral
BrandKeithley
ModelS530
CategoryTest Equipment
LanguageEnglish

Related product manuals