8.6 Weighing with tare
8.6.1 Taring
Deposit weighing container. After successful stability check press the TARE
button. Zero display and indicator NE T appear.
The weight of the container is now internally saved.
Weigh the material, the net weight will be indicated.
The weight of the weighing container will be displayed as a minus number after
removing the weighing container.
The tare procedure can be repeated as many times as necessary, for example
with initial weighing of several components for a mix (add-on weighing). The limit
is reached when the taring range capacity (see type plate)is full.
Press the BG/NET-key to switch between gross and net weight.
To delete the tare value, remove load from weighing pan and press the TARE-
key.
The NET indicator turns off, the zero display shows.
8.6.2 Numerical input of tare (PRE-TARE)
The known dead weight of a weighing container can be tared off by entering its
weight as pre-tare deduction in order to ensure the net weight of the goods to be
weighed in subsequent weighings is always displayed.
Press TARE-key on unloaded weighing scale / zero display and the enabled
place will start flashing.