21 - 2 Defibrillator/Monitor Operator’s Manual
21.4 Exporting Configurations
1. Connect the USB flash memory to the equipment’s USB connector.
2. Switch the Mode Select knob to Monitor, Manual Defib or Pacer.
3. Press the Main Menu button on the front panel.
4. Enter the Configuration Main menu by either of the following ways:
◆ Select [Others >>] → [Configuration >>] → [View Config].
◆ Select [Others >>] → [Configuration >>] → enter the required password
5. Select [Config. Export].
21.5 Importing Configurations
It is not necessary to configure each equipment separately when installing several equipments with identical
configurations. A USB flash memory can by used to import configurations from one equipment to another.
1. Prepare a USB flash memory with desired configurations.
2. Connect the USB flash memory to the target equipment’s USB connector.
3. Switch the Mode Select knob to Monitor, Manual Defib or Pacer.
4. Press the Main Menu button on the front panel.
5. Enter the Configuration Main menu by either of the following ways:
◆ Select [Others >>] → [Configuration >>] → [View Config].
◆ Select [Others >>] → [Configuration >>] → enter the required password
6. Select [Config. Import]
21.6 Changing Configurations
You can change configurations either by changing settings in each configuration setup menu, or restoring to
factory defaults.
1. Switch the Mode Select knob to Monitor, Manual Defib or Pacer.
2. Press the Main Menu button on the front panel, and then select [Others >>] → [Configuration >>] →
enter the required password.
â—† Selecting [Factory Default] can restore all the current settings to factory defaults.
â—† Select the desired setup menu, and change the settings.
â—† Selecting [Record] can record all the current settings.
21.6.1 General Setup Menu
Menu Item Options/Range Default Description
Device Name 20 characters / Inputs the relevant information.
The characters are included in
the keyboard. Restoring factory
default configurations does not
change these settings.
If the equipment is connected to
the CMS, [Bed No.] and
[Department]can also be
changed on the CMS.
Institution 20 characters /
Department 20 characters /
Bed No. 20 characters /
Patient Cat. Adu, Ped, Neo Adu Sets the default patient category.
Patient Info Display Bed No., Patient Name Bed No. Sets the patient information type
displayed in the upper left corner
of the main screen.
Height Unit cm, inch cm Sets the default measurement
unit for the patient.
Weight Unit kg, lb kg