Defibrillator/Monitor Operator’s Manual F - 1
F Electrical Safety Inspection
The following electrical safety tests are recommended as part of a comprehensive preventive maintenance
program. They are a proven means of detecting abnormalities that, if undetected, could prove dangerous to
either the patient or the operator. Additional tests may be required according to local regulations.
All tests can be performed by using commercially available safety analyzer test equipment. These procedures
assume the use of a 601PROXL International Safety Analyzer or equivalent safety analyzer. Other popular testers
complying with IEC 60601-1 used in Europe, such as Fluke, Metron or Gerb, may require modifications to the
procedure. Please follow the instructions of the analyzer manufacturer.
The electrical safety inspection should be periodically performed per year. The safety analyzer also proves to be
an excellent troubleshooting tool to detect abnormalities of line voltage and grounding, as well as total current
loads.
F.1 Power Cord Plug
F.2 Device Enclosure and Accessories
F.2.1 Visual Inspection
F.2.2 Contextual Inspection
Test Item Acceptance Criteria
The power plug The power plug pins No broken or bent pin. No discolored pins.
The plug body No physical damage to the plug body.
The strain relief No physical damage to the strain relief. No plug warmth for device in
use.
The power plug No loose connections.
The power cord No physical damage to the cord. No deterioration to the cord.
For devices with detachable power cords, inspect the connection at
the device.
For devices with non-detachable power cords, inspect the strain relief
at the device.
Test Item Acceptance Criteria
The enclosure and accessories No physical damage to the enclosure and accessories.
No physical damage to meters, switches, connectors, etc.
No residue of fluid spillage (e.g., water, coffee, chemicals, etc.).
No loose or missing parts (e.g., knobs, dials, terminals, etc.).
Test Item Acceptance Criteria
The enclosure and accessories No unusual noises (e.g., a rattle inside the case).
No unusual smells (e.g., burning or smoky smells, particularly from
ventilation holes).
No taped notes that may suggest device deficiencies or operator
concerns.

Table of Contents

Other manuals for Mindray BeneHeart D3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D3 and is the answer not in the manual?

Mindray BeneHeart D3 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart D3
CategoryMedical Equipment
LanguageEnglish

Related product manuals