21 - 14 Defibrillator/Monitor Operator’s Manual
21.6.19 Others Menu
Default Network Connection On, Off Off Sets whether the CMS is
automatically connected after
turned on.
Restoring factory default
configurations does not change
this setting.
Site Setup Select Site 1, 2 1 Sets the desired CMS.
Restoring factory default
configurations does not change
this setting.
Site Name 0 to 20 characters /
Site Address 0 to 63 characters /
HL7
Configuration
- Data +
Waveforms
Destination
Site
0 to 63 characters / Sets the desired HL7 server.
Port 0-65535 0
Send Data On, Off Off Sets whether monitoring
parameters are automatically
sent to the HL7 server.
Data Interval 1s, 30s, 5min, 30min, 1h 1 s Sets the interval for sending
data to the HL7 server.
Send
Waveforms
On, Off Off Sets whether waveforms are
automatically sent to the HL7
server.
HL7
Configuration
- Alarms
Destination
Site
0 to 63 characters / Sets the desired HL7 server.
Port 0-65535 0
Send Alarms On, Off Off Sets whether alarms are
automatically sent to the HL7
server.
Menu Item Options/Range Default Description
Menu Item Options/Range Default Description
Brightness 1 to 10 8 Sets the display brightness.
Key Volume 0 to 10 2 Sets the key volume.
Wave Line Thick, Medium, Thin Medium Sets the thickness of displayed
waveform.
Draw Wave Mono, Color Color Sets the color mode of
displayed waveform.
Modules SpO2 On, Off On Sets whether to enable the
parameter module.
NIBP
CO2
Resp
Modify Password 0 to 32 characters / The characters are included in
the keyboard. Restoring factory
default configurations does not
change this setting.