Defibrillator/Monitor Operator’s Manual 21 - 3
21.6.2 Manual Defib Setup Menu
Language /-- / Sets the language for voice and
text prompts.
Date Format yyyy-mm-dd, mm-dd-yyyy,
dd-mm-yyyy
yyyy-mm-dd Sets the system date format.
Time Format 12 h, 24 h 24 h Sets the system time format.
System Time Year 2007 to 2099 2007 Sets the system date.
Configurable range: 2007-01-01
to 2099-05-31.
Month 01 to 12 01
Day 01 to 31 01
Hour 24 h: 00 to 23
12 h: 12AM to 11PM
24 h: 00
12 h: 12AM
Sets the system time.
Minute 00 to 59 00
Second 00 to 59 00
Menu Item Options/Range Default Description
Menu Item Options/Range Default Description
Manual Therapy Access Direct, Confirmed, Password Direct Sets the way you access the
Manual Defib mode and Pacer
mode.
•[Direct]: you can directly
access the Manual Defib
mode and Pacer mode.
•[Confirmed]: a dialog box
pops up for the confirmation
when you access the Manual
Defib mode and Pacer mode.
•[Password]: a dialog box
pops up requiring the
password when you access
the Manual Defib mode and
Pacer mode.
Set Password 4 digits 0000 [Set Password] is active only
when [Manual Therapy Access]
is set to [Password].
Default Energy for Adult 100 J, 150 J, 170 J, 200 J, 300 J,
360 J
200 J Sets the energy level for the
external defibrillation.
Default Energy for Pediatric 10 J, 15 J, 20 J, 30 J, 50 J, 70 J,
100 J
50 J
Internal Default 2, 5, 10, 20, 30, 50 10 J Sets the energy level for the
internal defibrillation.
Sync After Shock Yes, No No Sets whether the equipment
remains in synchronized
cardioversion after a delivered
shock.
Remote Sync On, Off Off Sets whether to enable the
remote synchronized
cardioversion.
Time to Auto Disarm 30 s, 60 s, 90 s, 120 s 60 s Sets the time the equipment
automatically removes the stored
energy internally.
Monitor Para. SpO2, NIBP, CO2, Off Off Sets the parameter monitored in
the auxiliary parameter area of
the Manual Defib mode and AED
mode.