Defibrillator/Monitor Operator’s Manual A - 11
*Measurement accuracy verification: In adult and pediatric modes, the blood pressure measurements measured
with this device are in compliance with the Standard for Non-invasive sphygmomanometers (ISO 81060-2)in
terms of mean error and stardard deviation by comparing with intra-arterial or auscultatory measurements
(depending on the configuration) in a typical patient population. For auscultatory reference, the 5th Korotkoff
sound was used to determine the diastolic pressure.
In neonatal mode, the blood pressure measurements measured with this device are in compliance with the
American National Standard for Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and
stardard deviation by comparing with intra-arterial measurements (depending on the configuration) in a typical
patient population.
A.5.7 CO
2
Specifications
Auto mode repetition intervals 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 60, 90, 120, 180, 240 or 480 min
STAT mode cycle time 5 min
Static pressure measurement range 0mmHg to 300mmHg
Static pressure measurement accuracy ±3mmHg
Maximum measurement time Adult, Pediatric:
Neonate:
180s
90s
Initial cuff inflation pressure range Adult:
Pediatric:
Neonate:
80 to 280 mmHg
80 to 210 mmHg
60 to 140 mmHg
Default Initial cuff inflation pressure Adult:
Pediatric:
Neonate:
160 mmHg
140 mmHg
90 mmHg
Measurement range Adult Pediatric Neonate
Systolic mmHg 25 to 290 25 to 240 25 to 140
Diastolic mmHg 10 to 250 10 to 200 10 to 115
Mean mmHg 15 to 260 15 to 215 15 to 125
Software overpressure protection Adult: 297±3 mmHg
Pediatric: 297±3 mmHg
Neonate: 147±3 mmHg
Measurement accuracy* Max mean error: ±5 mmHg
Max standard deviation: 8 mmHg
Resolution 1 mmHg
Measurement range 0 to 150 mmHg
Accuracy* Full accuracy mode:
0 to 40 mmHg:
41 to 76 mmHg:
77 to 99 mmHg:
100 to 150 mmHg:
±2 mmHg
±5% of reading
±10% of reading
±(3 mmHg+8% of reading)
ISO accuracy mode: Add ±2mmHg to the full accuracy mode
Sart-up time 20 s (typical), 90 s (maximum)
Accuracy drift Meets the requirement for measurement accuracy within 6 hours.
Resolution 1mmHg
Sample flowrate Connecting the Oridion sampling line: 50 ml/min

Table of Contents

Other manuals for Mindray BeneHeart D3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart D3 and is the answer not in the manual?

Mindray BeneHeart D3 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart D3
CategoryMedical Equipment
LanguageEnglish

Related product manuals