Care and Maintenance 12-11
NO.
Item Method Tool
mechanism of
the monitor
2. Manually operate the monitor to make sure the
monitor can move normally when it is turned left/
right, lifted/ lowered, and no abnormal sound
exists.
none
3. Manually turn the monitor supporting arm, the
monitor can move up or down. When rotate the
arm to the
position, the monitor will be locked
in the middle position and can’t move left or right.
Make sure there is no obvious looseness.
none
4. Remove the rear cover of the control panel and
neck cover of the monitor, check by a wrench to
confirm that the fixing screw are free of looseness;
Visually check to confirm that the cables are not
scratched or clipped out that the core can be
seen.
screwdriver,
inner hexagonal
wrench 5
6.
Turning and
lifting
mechanism
1. Hold the lifting arm control handle, make sure that
the lifting arm can go up and down normally
without abnormal sounds or phenomena
2. Hold the handle to make sure that the control
panel can turn smoothly without abnormal sounds
or phenomena.
none
7.
Other mechanics
Check to confirm that there is no looseness to other
mechanical parts, no crack to cover and no conductive
parts show in sight.
none
12.3.5 Electrical Safety Inspection
Only technical professionals or engineers after training can perform electric safety inspection.
Please refer to appendix A: Electrical Safety Inspection for details.