Probes and Biopsy 13-17
When exit the Biopsy Verify menu without saving the setting, system will confirm
with you “Data have changed. Do you want to save the changes?” Click [OK] to
save the setting and return to Biopsy menu.
6. Scan to locate the target. Center the target in the electronic guidezone path.
7. Direct the needle into the area of interest for specimen.
8. After extraction of the biopsy sample is completed, gently remove the probe from the
body. To exit Biopsy menu: press <Esc>; or, navigate the cursor onto the Other menu.
9. Disassemble the items and properly dispose of these items as required.
DANGER:
Fail to match the guidezone displayed to the guide may cause
the needle to track a path outside the zone.
It is extremely important that when using the adjustable angle
biopsy guides, the angle displayed on the screen matches the
angle set on the guide, otherwise the needle will not follow the
displayed guidezone which could result in repeated biopsies
or patient injury.