Image Optimization 5-73
5.13 Image Preset
In the Image Preset interface, you can preset image parameters of exam modes and soft
menu and menu layout of imaging modes.
5.13.1 Image Preset
The image preset is used to set image parameters for all the probes and a specific probe
in a specific exam mode. The setting methods are similar although parameter items are
different in each image mode.
An example is provided below to demonstrate the steps of preset as in Adult ABD exam,
B/THI mode for the L7-3s probe, for special imaging modes (Free Xros M, 3D/4D, iScape),
please refer to the relevant sections of each mode:
1. Press <Setup> on the keyboard to enter the Setup menu.
2. Select “[Image Preset] [B/THI]” to enter the B mode parameters preset, as shown
in the following figure.
3. Select the exam mode and probe.
Select “Adult ABD” from the “Exam Mode” drop-down list.
Select “L7-3s” from the “Probe” drop-down list.
4. Image parameter preset
The [Adult ABD-All Probe] field on the left side of the screen displays the
parameter preset for all probes in Adult ABD exam mode.