For further information about device settings please see device online help on the “Settingsscreen
and its submenus.
3.5.4 Hardware Tests
The main device functions can be tested with the Functional Checks” option in About Device”.
The device self-test examines several device properties such as internal power supply, codec
function, and memory integrity. If a device property is correctly working, test status is
indicated as “OK”. If not all device properties are tested successfully, please contact your distributor
or Natus Sales representative. To perform the device self- test, simply touch the device self- test
button. The test will start automatically.
The Cable Test checks the ATA Cable and Patient Cable functionality. Please use the ALGO
7i Check Kit to test the Impedance levels with your cable. To do so, plug in your Patient Cable
in the black socket of your ALGO 7i device and attach all three clips to the metal bar of your Check Kit.
Then plug in your ATA Cable in the grey socket and connect to the left and right opening of the Check
Kit. Tap the Cable Check button and follow the on-screen instructions. Please note: Please do not
run this test without connecting cables to the ALGO 7i Check Kit. If your ATA Cable fails the Cable Check,
Screening will be disabled with this particular cable until a successful Cable Check has been passed.
If anything other than those values listed for each test step are displayed, please contact Natus
Technical Service or your Natus authorized service representative.
It might happen that one of the error messages shown in Table 3 occurs. Please follow the
recommended actions for troubleshooting mentioned in Table 3.
Error message
Recommended actions for troubleshooting
No transducer
found
Check if the ATA Cable is properly connected to the device.
If not, connect the ATA Cable to the device.
Test Result
“FAIL”
Check if the ATA Cable is placed correctly and tightly in the test cavity.
If not, place the ATA Cable hooks in the cavities.
Table 3: Probe test error messages and recommended actions