5 Cleaning
Cleaning the device and its accessories is very important for compliance with hygienic
requirements and to avoid any cross-infection. Please always consider local regulations and
read this section carefully.
Before cleaning the device, the device must be switched off and removed from all connected
components (e.g. power supply unit).
Wipe the surface of the device with a cloth slightly dampened with mild detergent or normal
hospital bactericides or antiseptic solution. The following quantities of chemical substances
are allowed to clean the device:
ethanol: 70-80%,
propanol: 70-80%,
aldehyde: 2-4%.
Approved cleaning solutions for cleaning ATA Cables are 70% isopropyl alcohol and mild
soapy water solution. The ATA Cable transducer body contains sensitive components. Never
attempt to clean the sound apertures on the transducer body with cleaning solution or by inserting an
object in the sound apertures.
Do not immerse the device into liquid and ensure that no liquid gets inside the device. Dry the device
with a lint-free cloth after cleaning.
Disposable accessories (Jelly Tab sensors and Flexicouplers) must be replaced between patients to
avoid cross-infection.
It is recommended that parts which can have direct contact with the patient are subject to standard
disinfecting procedures between patients. This includes physical cleaning and use of recognized
disinfectants.
When using a cleaning agent, please refer to the manufacturer's data sheet of the cleaning agent for
the minimum time period in which the wipe has to be in direct contact with the surface of the device
or accessory to ensure effectiveness of cleaning.
The device and its accessories are provided non-sterile and are not intended to be sterilized.