14 Cleaning, Disinfection and Maintenance
64 / 85 Instruction Manual Pentacam® AXL (G/70100/EN 1121 Rev04)
14.2 Disinfection
Recommendation: Use disinfection wipes suitable for medical
devices, for example:
Mikrozid sensitive wipes premium; Fa. Schülke & Mayr
Softpack 48 Stück / Art. Nr. 165711
Schülke & Mayr GmbH; Tel: +4940521000 / Fax: +494052100318
E-Mail@schuelke.com; www.schuelke.com
Note
Equipment damage caused by disinfectant solution
The disinfectant solution may damage the surface of the device if it is
sprayed directly on it.
Spray the disinfectant solution onto a cleaning cloth, do not spray it
directly on the device.
Disinfect the forehead rest after each examination.
If you do not use paper for the chinrest, disinfect the chinrest after
each examination.