PDR-7000 Operation Manual
7.5 Standard Programs
The PDR-7000 has ten programs of Fixed Program(A,B) and User Program(A-H),
which can be programmed by the user.
The following standard programs have already been written in (Fixed) Programs A
and B as factory setting.
○ Fixed Program A
The final prescription is obtained after the monocular full correction (subjective data) is store.
The binocular balance test and auto adjustment are performed in the final field. Then the
precise addition power is measured with the cross grid for near vision in the subjective field.
(R: right eye, L: left eye, B: binocular eye)
1. AR data entry Enter the AR data.
<Unaided field>
2. Unaided VA test Measure the unaided VA.
<Aided field>
3. LM data entry Enter the LM data.
4. Aided VA entry Measure the aided VA.
<Subjective field>
5. R: AR data check
6. R: Red-green test
7. R: Cross cylinder test (axis)
8. R: Cross cylinder test (power).
9. R: Red-green test
10. R: Best possible VA check
(monocular full correction)
11. L: AR data check
12. L: Red-green test
13. L: Cross cylinder test (axis)
14. L: Cross cylinder test (power)
15. L: Red-green test
16. L: Best possible VA check
(monocular full correction)
Check that the VA is about 0.7 in the AR data
Spherical refinement with the red-green chart (fog + 0.5)
Measure the cylindrical axis with the cross cylinder lens.
Measure the cylindrical power with the cross cylinder lens
Spherical refinement with the red-green chart (fog + 0.5)
Best possible VA on the chart with most positive power
Check that the VA is about 0.7 in the AR data.
Spherical refinement with the red-green chart (fog + 0.5)
Measure the cylindrical axis with the cross cylinder lens.
Measure the cylindrical power with the cross cylinder lens.
Spherical refinement with the red-green chart (fog + 0.5)
Best possible VA on the chart with most positive power
<Final field>
17. B: Binocular balance test
18. B: Stereo test
Test the binocular balance.
Check the steroscopic vision and enter the stereoparallax
- 96 -