Display key – changes type of
and after drying process.
Setup key – enter to main menu.
Test menu key –
function key for selecting drying
modes.
6. START-UP
On plugging to mains the moisture analyzer carries out a display test (all pictograms
and digits appear on the display). Then the display shows software name and
number, and enters weighing mode.
6.1. Moisture analyzer temperature stabilization period
Before start of measuring processes, it is necessary to wait until the moisture
analyzer is thermally stabilized. It is a period of so called moisture analyzer’s
self-heating. For moisture analyzer MAC series which before plugging to mains
was stored in room temperature, self-heating period takes approximately 30
minutes. For moisture analyzers that were stored in much lower temperatures
before plugging to mains (e.g. during winter period) thermal stabilization should
last approximately 4 hours. During self-heating period the indications on
moisture analyzer’s display may change. Correct operation of a moisture
analyzer is possible within temperature range specified in the technical
parameters, see point 1. It is recommended that ambient temperature changes
at moisture analyzer’s place of use are very small (slow).