18. ERROR MESSAGES
If moisture balance software detects an error a proper error message appears on
the graphic display. The error message contains number of error which informs
about its source.
Start mass exceeds permissible range (it refers to start
and adjustment – start mass)
Low limit of converter range exceeded
High limit of converter range exceeded
Value of mass or temperature exceeds permissible range
(it refers to mass or temperature adjustment – final mass
Operation time exceeded (tarring or printing caused by
lack of stability status in set time)
Inserted value is out of range (reg. setting reference
mass, date, etc.)
Error 1.3 AD Th H
The value of A/D divisions of the thermometer in the
weighing chamber
= 0 temperature sensor defected, heating of weighing
chamber shall not start, return to weighing mode by
pressing ESC key.
The value of A/D divisions of the thermometer in the
weighing chamber
= 1024 temperature sensor defected heating of weighing
chamber shall not start, return to weighing mode by
The value of A/D divisions of actuator thermometer = 0
temperature sensor defected, return to weighing mode by
pressing ESC key.
The value of A/D divisions of actuator thermometer =
1024 temperature sensor defected, return to weighing
mode by pressing ESC key.