1. TECHNICAL DATA
Type
MAC 50/1/NH
MAC 50/1/WH
MAC 50/1/NP.
MAC 50/1/NS
MAC 50/NH
MAC 50/WH
MAC 50/NP.
MAC 50/NS
MAC 110/NH
MAC 110/WH
MAC 110/NP.
MAC 110/NS
MAC 210/NH
MAC 210/WH
MAC 210/NP.
MAC 210/NS
Max capacity
50 g
50 g
110 g
210 g
Readability
0,1 mg
1 mg
1 mg
1 mg
Tare range
- 50 g
- 50 g
- 110 g
- 210 g
Max sample
mass
50 g
50 g
110 g
210 g
Readability of
moisture
readout
0,0001% 0,001 %
Repeatability
of moisture
content
reading
+/-0,24% (sample mass up to 2g),
+/-0,06% (sample mass up to 2-10g),
+/-0,04% (sample mass over 10g)
Drying
temperature
range
max. 160 °C (model NH, NP and NS)
max. 250 °C (model WH)
Heating
module
halogen (model NH and WH)
infrared emitter (model NP)
heater in metal cover (model NS)
Drying modes
4 drying modes: standard, quick, step, mild
Finish mode
3 modes: automatic, time defined, manual
Working
temperature
+10 °C - +40 °C
Power supply
230V 50Hz AC / 400W
Display
LCD (backlit)
Pan size
Φ 90 mm, h = 8 mm
Drying
chamber
dimensions
120 x 120 x 20 mm
Net weight /
Gross weight
5,1 / 7,2 kg
Packaging
dimensions
455 x 315 x 302 mm