XCell
TM
ATF System with C410:V4B Controller User Guide
5
1.2 XCell™ ATF System Control Scope and Objectives
The C410 v4 Controller provides the process control functionality of the stainless steel
XCell™ ATF 4 System, XCell™ ATF 6 System, XCell™ ATF 10 System and the XCell™ ATF 6
and 10 Single-use devices. The XCell™ ATF C410v4B controller is designed to:
1. Control ATF flow rates in both pressure and exhaust cycles
2. Provide a user interface capability for XCell™ ATF System control and monitoring
3. Set-up and select operational parameters
4. Display real-time process data and Alarms for error conditions
5. Provide “Batch control” and user hierarchy
6. Have PLC software upgrades in the field by replacing memory modules
7. Have 3 major enclosures, an Electronics Box (E-Box) a Pneumatics Box (P-Box) and a
Power Box
8. Allow the E-Box to operate an XCell™ ATF 4, 6 and 10 systems and XCell™ ATF Single-use
6 and 10 devices using custom software
9. Allow the P-Box hardware and process parameters to be specific to the size of the
particular XCell™ ATF System being controlled.
Control principles are detailed in Section 6.
2.0 Utility Requirements
UTILITY
REQUIREMENT
ADDITIONAL CONSIDERATIONS
PRIMARY
COMPRESSED AIR
Maximum 90 psig /
6.1 bar
Oil free, dry, filtered gas, i.e., medical
grade air
Typically: 35 psi / 2.4bar,
Regulated oil free, dry, filtered air
VACUUM SERVICE
Minimum -12.5 psig
Vacuum supplied by a Repligen or
customer supplied local pump capable of
maintaining ~-12.5 psig with nominal
flow as follows:
XCell™ ATF4 - 40L/min
XCell™ ATF6 – 60L/min
XCell™ ATF10 – 200L/min
Pump should be clean room compatible
Untreated discharge or user specified
ELECTRICAL
100-240V AC, 50/60Hz
NA
Temperature: 0-50
°
C
Relative Humidity: 0 80
% RH
NA
STEAM (SIP)
~30lbs/hr.
Applicable only to a steamable
connection between XCell™ ATF and
bioreactor vessel
For SIP condensate drainage

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN XCell ATF10 and is the answer not in the manual?

REPLIGEN XCell ATF10 Specifications

General IconGeneral
BrandREPLIGEN
ModelXCell ATF10
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals