RIGOL Chapter 1 Overview
1-2 RSA3000 Performance Verification Guide
Step: 1 dB
Attenuation range: 0 dB to 11 dB
Frequency range: 50 MHz ± 1 MHz
Connector: N (F)
Attenuation repeatability: ±0.03 dB
Step: 10 dB
Attenuation range: 0 dB to 110 dB
Frequency range: 50 MHz ± 1 MHz
Connector: N (F)
Attenuation repeatability: ±0.03 dB
Attenuator
Interconnect Kit
Connect Agilent 8494H and Agilent 8496H
Impedance: 50 Ω, nominal
Connector: N (M)
Note
[1]
: The connecting terminals of the low-pass filter used in this manual are two SMA female terminals. In the
actual test, the type of the connecting terminals of the low-pass filter may differ; please select proper connecting
cable and adaptor according to the actual type of the connecting terminals.
Test Result Record
Record and keep the test result of each test. In the Appendix of this manual, a test result record form
which lists all the test items and their corresponding performance limits as well as spaces for users to
record the test results, is provided.
Tip:
It is recommended that users photocopy the test record form before each test and record the test
results in the copy so that the form can be used repeatedly.
Specifications
The specification of each test item is provided in Chapter 2. For other technical parameters, refer to
RSA3000 DataSheet
(available to download them from RIGOL website: www.rigol.com).
Tip:
All the specifications are only valid when the spectrum analyzer has been warmed up for more than
30 minutes.