Displayed Average Noise Level (DANL) Test
Specification
Displayed Average Noise Level (DANL)
RSA3015N
RSA3030
RSA3030N
RSA3045
RSA3045N
attenuation = 0 dB, sample detector, trace averages ≥ 50, tracking generator
off, normalized to 1 Hz, 20℃ to 30℃, input impedance = 50 Ω.
Preamp
off
9 kHz to 100 kHz
<-120 dBm (typical)
<-120 dBm (typical)
100 kHz to 20 MHz
<-135 dBm, <-140 dBm (typical)
<-135 dBm, <-140 dBm
(typical)
20 MHz to 2.7 GHz
<-138 dBm, <-141 dBm (typical)
<-138 dBm, <-141 dBm
(typical)
2.7 GHz to 3.0 GHz
<-136 dBm, <-141 dBm (typical)
<-136 dBm, <-141 dBm
(typical)
3.0 GHz to 4.5 GHz
<-136 dBm, <-140 dBm
(typical)
Preamp
on
100 kHz to 20 MHz
<-152 dBm, <-160 dBm (typical)
<-152 dBm, <-160 dBm
(typical)
20 MHz to 2.7 GHz
<-158 dBm, <-161 dBm (typical)
<-158 dBm, <-161 dBm
(typical)
2.7 GHz to 3.0GHz
<-156 dBm, <-161 dBm (typical)
<-156 dBm, <-161 dBm
(typical)
3.0 GHz to 4.5 GHz
<-154 dBm, <-159 dBm
(typical)
Test Device
1. 50 Ω Matched Load × 1
Test Connection Diagram
RSA3000
Load
RF IN
Figure 2-1 DANL Test Connection Diagram

Other manuals for Rigol RSA3030

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol RSA3030 and is the answer not in the manual?

Rigol RSA3030 Specifications

General IconGeneral
BrandRigol
ModelRSA3030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals