Chapter 2 Performance Verification Test RIGOL
RSA3000 Performance Verification Guide
2-15
TOI Distortion Test
Specification
TOI Distortion
fc ≥ 50 MHz, two -20 dBm tones at input mixer spaced by 200 kHz, attenuation = 0 dB, preamp off.
TOI
+10 dBm, +15 dBm (typical)
Test Devices
1. Signal Generator × 2
2. Power Meter × 1
3. Power Sensor × 1
4. Power Meter Connecting Cable × 1
5. Power Divider × 1
6. 300 MHz Low-pass Filter × 1
7. 1 GHz Low-pass Filter × 1
8. Dual-N Cable × 2
9. N-SMA Cable × 2
10. Dual-BNC Cable × 1