Setup and Device Conguration
Stryker Endoscopy considers instructional training an integral part of the CrossFlow system� Your Stryker
Endoscopy representative will perform at least one in-service at your convenience to help you set up your
equipment and instruct you and your sta on its operation and maintenance� Please contact your local
Stryker Endoscopy representative to schedule an in-service after your equipment has arrived�
1� Choose a location for the CrossFlow pump�
• Place the pump on a Stryker cart or other sturdy platform near a hospital grade outlet�
• Place the pump at the same height as the joint to ensure accurate pressure readings.
• Provide at least four to six inches of space around the sides of the pump to ensure proper ventilation
and allow access to the power cord�
Warning
RF and other mobile communications equipment may aect the normal function of the
CrossFlow pump� When placing the pump, follow the instructions located in the “Electromagnetic
Compatibililty” section of this manual�
2� Connect the AC power�
• Connect the provided hospital power
cord to the AC inlet on the rear console
panel�
• Connect the other end to a hospital-grade
power outlet�
Warning
• Check the device label on the rear of the pump to determine the operating voltage of the
device�
• Check the power cord assembly periodically for damaged insulation or connectors�
• To avoid risk of electric shock, this equipment must only be connected to a supply mains with
protective earth�
Caution
• When connecting or disconnecting a cable, hold the cable by its connector (its plug, not the
cord)� Failure to comply may result in damage to the cable or pump�
• Connect the power cords directly to the AC inlet or outlet� Do not connect any of the power
cords together�
If required, connect the pump to a Stryker Firewire-compatible device using one of the SFB connector