• Empty the measuring cup�
• Unclamp the tubing and place it in the cup� Start the stopwatch once the tubing is placed back in
the cup�
• Once the measuring cup is lled with 1 liter of uid, press
to stop the pump�
• The test has been successfully completed if the time it takes to ll the measuring cup with 1 liter of
uid is within the range specied in the table below�
Flow (%) Time (seconds)
50 70
Perform Preventive Maintenance and Calibration
Caution
Stryker does not accept any liability for direct or consequential damages if:
• the pump or the accessories are improperly prepared and maintained,
• non-authorized persons perform repairs, adjustments, or alterations to the pump or accessories,
• non-authorized persons open the pump,
• the prescribed inspection and maintenance schedules are not followed�
When the pump is 95% through the current maintenance period, a “Service Pump Soon” notication
will appear on the screen� An authorized Stryker service technician must inspect and service the device
according to the maintenance and calibration schedule below to maintain product functionality�
Component Maintenance/Calibration Period
Motors 2years
Pinch valves 2years
Pressure transducer 2years
Expected Life
Equipment Expected Life
Console Fiveyears
Footswitch Threeyears
Hand Control Oneyear
Inow, outow, patient-use tubing Single-use
Day-use cassette Tencases, eighthours of active use, or 24hours after point of rst
use
Disposal
This product contains electrical waste or electronic equipment� It must not be disposed of as unsorted
municipal waste and must be collected separately in accordance with applicable national or institutional
related policies relating to obsolete electronic equipment�
Dispose of any system accessories and irrigation uid according to normal institutional practice relating to
potentially contaminated items�