EN-27
DRAFT
Navigating to the Help Menu
1� Press
to open the Help Menu on any screen�
System Feedback
Audible Feedback
Audible Feedback Event Implication
Three high-tone beeps Fault error Refer to “Error Messages”
Three high-tone beeps,
pause, two beeps
Lockdown error Refer to “Error Messages”
One medium-tone
beep
Accessory connection Refer to “Connect the Components” and “Insert
the Inow/Outow Cassette”
Two medium-tone
beeps
Accessory removal Refer to “Connect the Components” and “Insert
the Inow/Outow Cassette”
One low-tone beep Touchscreen/Hand control/
Footswitch operation
N/A
Three (medium, low,
high-tone) beeps
Hot swap Refer to “Swap between Hardware Combination
Selections”
Error Messages
The words caution, fault, and lockdown carry special meanings and should be carefully reviewed:
• Caution: Visual popup error is displayed for ve seconds; user may clear the message when the problem
is resolved�
• Fault: The pump will stop; user may clear the message when the problem is resolved�
• Lockdown: The pump will stop; reboot system�
Error Message Cause Possible Solution
CAUTION Cassette about to expire (two hours left
before expiration)�
• Replace the cassette�
CAUTION Preventative Maintenance required; the
unit has reached its recommended service
interval�
• Contact your Stryker
representative�
CAUTION Inow/Outow RFID read failure: The
Inow or Outow RFID tag cannot be
detected while the pump is stopped�
• Ensure the Inow or Outow
Cassette is fully inserted�
• If the problem persists, replace
the cassette�
• If the problem persists, contact
your Stryker representative�
FAULT Priming Error: No uid in the tubing after
ten seconds�
• Ensure the pinch clamps on the
saline bags are open�
• Ensure the Inow Cassette is
fully inserted�