Performance
Check-213
Service
SEC/DIV
Time-Mark
Setting
Generator
Setting
1 µs
.5m
0.2
ms
1 µs
.5 m
0.1
ms
TABLE 5-3
Sweep
Linearity
VOLTS-mA/DIV
Setting
(black scale)
.1
.5
.1
.5
HORIZ
MGF
VAR
Setting 1
CAL
detent
CAL
detent
Set
fully
clock-
wise.
Then
adjust
counterclockwise
to
1
marker/divi-
sion
Set
fully
clock-
wise.
Then
adjust
counterclockwise
to
1
marker/divi-
sion.
1
The
horizontal POS control may need adjustment
to
align a timing
mark
with
a vertical graticule line.
c.
Return
HOR
IZ
MGF
VAR
to
the
CAL
detent.
d.
Disconnect
the
Generator.
4.
CHECK
INTERNAL
TRIGGER
a.
Set:
VOL
TS-mA/DIV
SEC/DIV
INPUT
COUPLING
TRIGGER
HORIZ
MGF
1
(black
scale)
.5 m
DC
INTL
AC
AUTO
CAL
detent
b.
Connect
the
probe
to
the
Sine-Wave
Generator
with
a
dual
banana-plug
to
BNC
adapter
and
a
probe
adapter.
Set
the
Generator
frequency
to
1
kHz
and
adjust
the
out-
put
amplitude
for
a 0.5 division
display.
c. CH
ECK-Sweep
triggers
at
some
point
on
both
the
positive
and
negative
slope
of
the
LEVEL-SLOPE
con-
trol,
and
free-runs
when
not
triggered.
d.
Set
TRIGGER
to
INTL
AC
NORM.
5-6
e.
CHECK-Sweep
triggers
at
some
point
on
both
the
positive
and
negative
slope
of
the
LEVEL-SLOPE
con-
trol,
and
there
is
no
sweep
when
not
triggered.
f.
Set
TRIGGER
to
INTL
AC
AUTO
and
SEC/DIV
to
2 µ.
g.
Set
the
Generator
to
1 MHz
with
an
output
ampli-
tude
of
0.5 division.
h.
Repeat
steps
c
through
e.
i.
Disconnect
the
Generator.
5.
CHECK
EXTERNAL
TRIGGER
a.
Set:
VOLTS-mA/DIV
SEC/DIV
TRIGGER
.5
.5 m
INTL
AC
AUTO
b.
Connect
the
Sine-Wave
Generator
to
EXT
TRIG
and
the
probe
using
two
dual
banana-plug
to
BNC
adapters,
a
T-Connector,
a
probe
adapter,
and
a
cable.
Set
the
Generator
to
1
kHz
and
adjust
the
output
amplitude
for
a 2 division
display
( 1 V p-p).
c.
Set
TRIGGER
to
EXT
DC.
d.
CHECK-Sweep
will trigger
at
some
point
on
both
the
positive
and
negative
slope
of
the
LEVEL-SLOPE
control
(there
is
no
trace
when
not
triggered).
e.
Set
TRIGGER
to
INTL
AC
AUTO
and
SEC/DIV
to
2µ.
f.
Set
the
Generator
to
1 MHz
and
adjust
the
output
amplitude
for
a 2 division
display
(1
V p-p).
g.
Repeat
steps
c
and
d.
h.
Disconnect
the
Generator.
@