Calibration-213
Service
5. CHECK
EXTERNAL
TRIGGER
a.
Set:
VOL
TS-mA/DIV
SEC/DIV
TRIGGER
.5
.5 m
INTL
AC
AUTO
b.
Connect
the
Sine-Wave
Generator
to
EXT
TRIG
and
the
probe
using
two
dual
banana-plug
to
BNC
adapters,
a
T-Connector,
a
probe
adapter,
and
a
cable.
Set
the
Generator
to
1 kHz and
adjust
the
output
amplitude
for
a 2
division
display
(1
V p-p).
c.
Set
TRIGGER
to
EXT
DC.
6-12
d.
CHECK-Sweep
will trigger
at
some
point
on
both
the
positive
and
negative
slope
of
the
LEVEL-SLOPE
control
(there
is
no
trace
when
not
triggered).
e.
Set
TRIGGER
to
INTL
AC
AUTO
and
SEC/DIV
to
2µ.
f.
Set
the
Generator
to
1 MHz
and
adjust
the
output
amplitude
for
a 2 division
display
(1
V p-p).
g.
Repeat
steps
c
and
d.
h.
Disconnect
the
Generator.
@