EasyManuals Logo
Home>Tektronix>Test Equipment>2215A

Tektronix 2215A Operator's Instruction Manual

Tektronix 2215A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
Operating
Pro
ce
dur
es-
2215A Ope
ra
tors
Pu
lse Jitter Time Measurement
To
measure pulse jitter time:
1. Perform steps 1 through 7
of
the preceding "Magni-
fied
Sweep Runs After Delay· procedure.
2. Referring
to
Figure 4-11 , measure the difference be-
tween Point C and Point D
in
divisions and calculate the
pulse jitter time using the following formula:
Pulse Horizontal B SEC/DIV
Jitter difference x
sw
itch
Time (divisions) setting
Triggered Magnified Sweep
The following procedure explains
how
to
operate the B
Sweep
in
a triggered mode and
to
determine the resulting
apparent magnification factor. Operating the B Sweep
in
a
triggered mode provides a more stable display, since the
delayed display is triggered at the same poi
nt
each time.
1. Perform steps 1 through
7
of
the preceding "Magni-
fied
Sweep Runs After Delay" procedure.
2. Adjust the B TRIGGER LEVEL control
so
the intensi-
fied zone on the A trace is stable.
NOTE
The
intensified zone seen in the
ALT
HORIZONTAL
MODE
display will move from trigger
point
to
trigger
point
as the B DELAY TIME POSITION control
is
rotated.
3. The apparent magnification factor can be calculated
from the formula shown
in
step 8
of
the "Magnified Sweep
Runs After Delay· procedure.
DELAYED-SWEEP TIME MEASUREMENTS
Operating the 2215A Oscilloscope with HORIZONTAL
MODE set
to
either ALT
or
B will permit time measurements
to
be made with a greater degree
of
accuracy than attained
with HORIZONTAL MODE set to
A.
The following proce-
dures describe how these measurements are accomplished.
Time Di
ffe
ren
ce
On Single Waveform
To
measure time between
two
points on a waveform,
use the following procedure:
4-10
I
!.
....
I.
A TRAC E
r
B TRAC E
r
I
t \
··
I
I
INTENSIFIED ZONE
TO BE MAG
NI
FIED
........
~
~
. ,
,,,
.
.
..
.
·
·\
\
~
~I
'II
.
...
.... ....
M
~
I
I
I
I
--
. ... ..
,,
MEASURE
I
I
TI
ME
FROM-I
i-
JIT
TER
@
TO
@
4732·11
Figure 4-11. Pulse jitter.
1 . Preset instrument controls and obtain a baseline
trace.
2. Apply the signal
to
either vertical-channel input con-
nector and set the VERTICAL MODE switch
to
display the
channel used.
3. Set the appropriate VOLTS/DIV switch
to
produce a
display
of
approximately 2-
or
3-
divisions in amplitude.
4. Set the A SEC/DIV switch
to
display the measure-
ment points
of
interest on the waveform. Ensure that the
SEC
/DIV Variable control is in the CAL detent.
5. Select ALT HORIZONTAL MODE and adjust both the
appropriate vertical POSITION control and A/B SWP
SEP
control to display the A trace above the B trace (see Figure
4-12).
6.
Set the B SEC/DIV switch to the fastest sweep speed
that provides a useable (visible) intensified zone.
7. Adjust the 8 DELAY TIME POSITION control
to
move
the intensified zone
to
the leading edge on the first point
of
interest (on the A trace); then fine adjust until the selected
portion (on the B trace) is centered at any convenient verti-
cal graticule line (see Figure 4-12).
8. Record the B DELAY POSITION dial setting.

Table of Contents

Other manuals for Tektronix 2215A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2215A and is the answer not in the manual?

Tektronix 2215A Specifications

General IconGeneral
BrandTektronix
Model2215A
CategoryTest Equipment
LanguageEnglish

Related product manuals