Performance Check Procedure—2215A Service
HORIZONTAL
Equipment Required (see Table 4-1):
Calibration Generator (Item 1)
50-0 BNC Cable (Item 4)
Leveled Sine-Wave Generator (Item 2)
50-0 BNC Termination (Item 5)
Time-Mark Generator (Item 3)
INITIAL CONTROL SETTINGS
Vertical
POSITION (both)
VERTICAL MODE
BW LIMIT
CH 1 VOLTS/DIV
CH 1 VOLTS/DIV
Variable
Channel 1 Input Coupling
Midrange
CH 1
Off (button out)
0.5 V
CAL detent
DC
Horizontal
POSITION Midrange
HORIZONTAL MODE A
A SEC/DIV 0.05 #xs
SEC/DIV Variable CAL detent
X I0 Magnifier Off (knob in)
B DELAY TIME POSITION Fully counterclockwise
B TRIGGER
SLOPE OUT
LEVEL Fully clockwise
A TRIGGER
VAR HOLDOFF NORM
Mode NORM
SLOPE OUT
LEVEL Midrange
A&B INT VERT MODE
A SOURCE INT
A EXT COUPLING DC -- 10
PROCEDURE STEPS
1. Check Timing Accuracy and Linearity
a. Connect the time-mark generator output via a 50-0
cable and a 50-0 termination to the CH 1 OR X input
connector.
b. Select 50-ns time markers from the time-marker
generator.
c. Adjust the A TRIGGER LEVEL control for a stable,
triggered display.
d. Use the Horizontal POSITION control to align the sec
ond time marker with the second vertical graticule line.
e. CHECK—Timing accuracy is within 3% (0.24 division
at the 10th vertical graticule line), and linearity is within 7%
(0.14 division over any 2 of the center 8 divisions).
4-7