EasyManuals Logo
Home>Tektronix>Test Equipment>314

Tektronix 314 User Manual

Tektronix 314
166 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #82 background imageLoading...
Page #82 background image
Performance Check/Calibration-314 Service
7. Adjust CH 1 Gain
Set controls as follows:
CH 1 and CH 2
VOLTS/DIV 10 m
VARIABLE CAL
lnput Coupling
CH 1 DC
CH
2
GND
DISPLAY CH 1
CHOP and ALT out
a. Connect Standard Amplitude Calibrator through
50-ohm cable to CH 1 and CH 2 VERT INPUT connectors
via Dual lnput Cable.
b. Set Standard Amplitude Calibrator for 50 mV
square-wave.
c. Check for 5 divisions of vertical deflection,
f3%
(0.75 minor division).
d. ADJUST-R43 (see Fig. 7-4), Gain, for exactly 5
divisions of vertical deflection. For SN 300422-up, switch
CH 1
VOLTS/DIV to 1 mV and standard amplitude
calibrator to 5 mV. Adjust R33B for exactly 5 divisions of
vertical deflection. Return standard amplitude calibrator
to 50 mV setting.
8.
Adjust CH
2
Gain
Set controls as follows:
CH 1 Input Coupling GND
CH
2
Input Coupling DC
DISPLAY CH 2
a. Check CH
2
for 5 divisions of vertical deflection,
f
3O/o, (0.75 minor division).
b. ADJUST-R83 (see Fig. 7-4), Gain, for exactly 5
divisions of vertical deflection. For SN 300422-up switch
CH 2
VOLTS/DIV to
1
mV and standard amplitude
calibrator to
5
mV. Adjust R73B for exactly 5 divisions of
vertical deflection.
c. Disconnect standard amplitude calibrator signal
from the 314.
9. Adjust and Check Volts/Div Compensation
a. Install calibration shield. See Fig. 5-2.
b. Connect Square-Wave Generator High-Amplitude
Output through 50-ohm cable, 50-ohm termination, and
47 pF Normalizer, to CH 1 VERT INPUT connector.
Fig.
5-2.
Calibration shield in place on
314.
between cable and termination, if necessary to maintain 5
divisions of vertical display.
e.
ADJUST-C1 (see Fig. 7-4) for flat top on displayed
square-wave. See Fig. 5-3 for typical display.
f. Using Table 5-6 (CH 1 Adjustments) check or adjust
for best flat top and front corner.
g. Move the signal (and Normalizer) to CH 2 VERT
INPUT connector.
ADJUST
FOR
F
LAT TOP
AND BOTTOM
OF SQUARE
WAVE
c. Set DISPLAY to CH 1, CH 1 lnput Coupling to DC,
TIME/DIV to 1 ms, CH
2
lnput Coupling to GND, and CH 1
&
CH 2 VOLTS/DIV to 10 m.
d. Set Generator to display about 5 cycles of square-
wave, using an amplitude sufficient to produce about 5
vertical divisions of display. Use a 50-ohm attenuator,
--
Fig.
5-3.
Typical display of correct adjustment of
C1
or
C11.
REV.
B.
AUG. 1976

Table of Contents

Other manuals for Tektronix 314

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 314 and is the answer not in the manual?

Tektronix 314 Specifications

General IconGeneral
BrandTektronix
Model314
CategoryTest Equipment
LanguageEnglish

Related product manuals