56
AFG1000 Series Quick Start User Manual
Appendix A: Line fuse replacement
The line fuse is in the plastic fuse box below the power line input on the rear panel.
WARNING. Disconnect the line cord at the rear panel and remove all test leads
connected to the instrument before replacing the line fuse. Failure to do so could
expose the operator to hazardous voltages that could result in personal injury or
death.
Use only the correct fuse type. Failure to do so could result in personal injury or
instrument damage.
Voltage
Fuse
100 - 120 V
250 V, F1AL
220 - 240 V
250 V, F0.5AL
To perform the line fuse replacement, follow these steps:
1.
Remove the power cord.
2.
Use a screwdriver to
remove the fuse box.
3.
Replace the fuse with a
new one, which should
match with the voltage;
install it into the fuse box
and then push the fuse
box back on to the rear
panel.